Starting up the Electron Microprobe
UHH / HVO Electron Microprobe Home Shutting down the Electron Microprobe Electron Microprobe Manual

TURNING ON THE HIGH VOLTAGE POWER AND CHECKING BEAM SETTINGS

** Move to the Brass Saturation Spot **

1) X-RAY Amplifier (Detector High Voltage)

2) Gun High Voltage

  • Make sure that the kV buttons on the SEM console are set to 0 kV.

  • Turn on the Electron Gun Power Switch

  • Incrementally raise the kV by depressing the 5, 10, and 15 kV buttons sequentially.

  • NEVER JUMP kV BUTTONS.

3) Beam Unblanked

  • Check that the Beam Blank Button is in the out position and not lit up brightly.

  • Do not use the physical beam blank button, use the button in the computer.

4) Raster Size

  • Make sure that the Raster Size knob is set to point.

5) Beam Size